Relative Maximum And Relative Minimum Hold - Unipulse F370 Operation Manual

Digital indicator
Table of Contents

Advertisement

7-5-5. Relative Maximum and Relative Minimum Hold

Detection starts when the indicated value crosses the auto start level with the H/M signal
inputted. Detection is performed as long as the H/M signal is on.
The hold is released by turning on the T/H signal as a reset signal.
The H/E output signal is on between the instant when the hold starts and the instant when
the T/H signal is turned on.
If the auto start level is not needed as a starting condition of the detection period, set
99999.
(Example) Relative Maximum Hold
Auto Start Level
H/M
T/H
HOLD
Buttun
H/E
HOLD
PEAK
t1: A delay time between the instant when the indicated value exceeds the auto start
level and the instant when the value to be held is detected
0.5ms (max.)
t2:A delay time between the instant when the H/M signal is inputted and the instant
when the value to be held is detected
1.0ms (max.)
t3:A minimum reset signal width required for releasing the hold
1.0ms (min.)
Indicated Value
t1
Detection
t2
OFF
ON
OFF
ON
OFF
ON
OFF
HI-LO Limit
Judging Output
ON
OFF
ON
OFF
ON
OFF
Blink
※ Comes on if a value is held with conditions met in the
detection period, and goes out if a value is not held.
7.HOLD FUNCTIONS
Sensor Input Value
Hold Period
t3
Decide
※ ON
or
OFF
OFF
ON
OFF
t
57