  Digital Zero conducted by External signal (D/Z input)
Digital Zero activates in the moment when the connection of the external input
terminal D/Z on the rear panel and COM2 is switched from Open to Short-circuit
then the indicated value becomes Zero.

5-2. Digital Filter

The digital filter is a function for reducing drifts of the indicated value by means of a
moving average of data converted from analog to digital. The number of the moving
averages can be selected a range between 0 and 512.
With an increase in the number of filterings, the indicated value becomes more stable,but
the response to inputs becomes slower.
How to set
Setting call
The value of a digital filter is scrolled with a
OFF
D/Z
ON
Keep on for 1.0 msec. or more.
CAUTION
In case of power interruption Digital Zero is released.
After recovering from power interruption, carry out the digital zero
again if necessary.
Number of settings:OFF, 2, 4, 8, 16, 32, 64, 128, 256, 512
Operation Setting
5.SETTING OF FUNCTIONS
* See the section on About a
Page 1
Setting Call page 8 .
button.
Scroll button