7-6-3. Time Specified Period Hold

(Peak, Valley, Peak-to-Peak and Mean Value)

Hold is detected during the predetermined time (hold detection time) from the point in
time when the H/M signal is turned on. The hold is released by turning on the T/H signal
as a reset signal.
The H/E output signal is on between the instant when the H/M signal is turned off and
the instant when the T/H signal is turned on.
(Example) Time Specified Period Peak Hold
H/M
HOLD
Button
HOLD
Time
T/H
H/E
HOLD
PEAK
t1:A delay time between the instant when the H/M signal is inputted and the instant
when the value to be held is detected
1.0ms (max.)
t2:A delay time between the instant when the hold detection time is expired and the
instant when the value to be held is determined
1.0ms (max.)
t3:A minimum reset signal width required for releasing the hold
1.0ms (min.)
Indicated Value
Detection
t1
OFF
ON
OFF
ON
OFF
ON
OFF
ON
OFF
HI-LO Limit
Judging Output
ON
OFF
ON
OFF
ON
OFF
Blink
7.HOLD FUNCTIONS
Sensor Input Value
Hold Period
t3
t2
Decide
ON
ON
t
OFF
OFF
63