Torque Settings; Calibration; Calibration Procedure For Torque; Utm Capacity (Utm Cap.) - Unipulse TM400 Operation Manual

Portable torque monitor
Table of Contents

Advertisement

4 Torque Settings
4

Torque Settings

4-1.

Calibration

"Calibration" is the operation of matching the TM400 with a sensor.
Check the UTMⅡ to be used.
4-2.

Calibration procedure for torque

Chapter
4
*

UTM capacity (UTM CAP.)

Select the rated capacity of UTMⅡ connected.

Other capacity (OTHER CAP.)

Set the arbitrary torque value against 5V input from UTMⅡ.

Unit (UNIT)

Set the unit of torque to display.
16
16
*
(1)
UTM capacity
*
(2)
Sign
*
(3)
Zero calibration
(4)
Minimum scale
:
Be sure to perform the setting.
<Setting range>
0.03, 0.05, 0.1, 0.2, 0.3, 0.5, 1, 2, 3, 5, 10, 20, 30, 50,100, 200,
300, 500, 1000, 2000, 3000, 5000, 10000 [Nm], OTHER
* The following setting items only available when OTHER is selected, and arbitrary calibration
can be made.
- Other capacity
- Unit

(These items only available in case the UTM capacity is set to OTHER)

<Setting range>
1 to 99999
(These items only available in case the UTM capacity is set to OTHER)
<Setting range>
mNm, Nm, kNm
Register the rated capacity value of the sensor.
Set the sign of the sensor.
Set the zero point of the sensor in an unloaded condition
(where no force and so on is applied to a sensor).
Set the minimum unit of torque such as scale interval
and scale division. (optional if there is no change)
- Decimal point