Hold Operation; Sample Hold - Unipulse F370 Operation Manual

Digital indicator
Table of Contents

Advertisement

7-5. Hold Operation

7-5-1. Sample Hold

When the T/H signal is inputted, a desired point is held, and the H/E output is turned on.
Hold of the value continues as long as the T/H signal is on.
T/H
HOLD
Button
HI-LO Limit
Judging Output
H/E
HOLD
Status Display
PEAK
Status Display
t1: A delay time between the instant when the T/H signal is inputted and the instant
when the indicated value is held
1.0ms (max.)
t2:A delay time between the instant when the T/H signal is released and the instant
when the indicated value returns to tracking
1.0ms (max.)
t3:A minimum reset signal width required for releasing the hold
1.0ms (min.)
Sensor Input Value
t1
OFF
ON
OFF
ON
OFF
ON
OFF
ON
Blink
Detection ・ Hold Period
Decide
ON
OFF
7.HOLD FUNCTIONS
Indicated Value
t2
t
t3
Blink
53