7-6. How to Specify the Hold Detection Period

7-6-1. All Period

By this method, the hold detection period is externally specified by the T/H signal.
Detection starts with the T/H signal ON to perform each hold operation.
Detection ends with the T/H signal OFF to reset the hold.
(Example) All Period Peak Hold
T/H
HOLD
Button
HI-LO Limit
Judging Output
H/E
HOLD
Status Display
PEAK
Status Display
t1: A delay time between the instant when the T/H signal is inputted and the instant
when the indicated value is held
1.0ms (max.)
t2: A delay time between the instant when the T/H signal is released and the instant
when the indicated value returns to tracking
1.0ms (max.)
t3:A minimum reset signal width required for releasing the hold
1.0ms (min.)
Sensor Input Value
t1
Detection
Hold Period
OFF
ON
OFF
ON
OFF
Undetermind
Period ( Note)
ON
OFF
ON
OFF
ON
OFF
Blink
Note: During the undetermined period, the judging output varies with
fluctuations in the input waveform. However, the H/E output
remains on during the undetermined period. Read the judging
result when the indicated value becomes stable (immediately
before the T/H signal rises).
7.HOLD FUNCTIONS
Indicated Value
t2
t1
t2
Detection
Hold Period
t3
t3
Undetermind
Period ( Note)
OFF
ON
OFF
OFF
Blink
OFF