Procedure
See
Figure
Figure 11-11:
D3 or D3 and D4 density calibration using ProLink III
downstream from sensor
D3 calibration
Fill sensor with D3 fluid
Device Tools >
Calibration >
Density Calibration >
Density Calibration – Point 3
Enter density of D3 fluid
11.6.3
Perform a D3 or D3 and D4 density calibration using the
Field Communicator
Prerequisites
Configuration and Use Manual
-
Minimum difference of 0.1 g/cm
density of the D3 fluid. The density of the D4 fluid must be greater than the
density of the D3 fluid.
-
Minimum difference of 0.1 g/cm
density of water. The density of the D4 fluid may be either greater or less than
the density of water.
Before performing the calibration, record your current calibration parameters. You
can do this by saving the current configuration to a file on the PC. If the calibration
fails, restore the known values.
11-11.
Close shutoff valve
Start Calibration
Close
During density calibration, the sensor must be completely filled with the calibration
fluid, and flow through the sensor must be at the lowest rate allowed by your
application. This is usually accomplished by closing the shutoff valve downstream
from the sensor, then filling the sensor with the appropriate fluid.
3
between the density of the D4 fluid and the
3
between the density of the D4 fluid and the
D4 calibration
Fill sensor with D4 fluid
Device Tools >
Calibration >
Density Calibration >
Density Calibration – Point 4
Enter density of D4 fluid
Start Calibration
Close
Done