Measurement support
11.2.1

Smart Meter Verification requirements

To use Smart Meter Verification, the transmitter must be paired with an enhanced core
processor, and the Smart Meter Verification option must be ordered for the transmitter.
See
Table 11-1
communication tool needed to support Smart Meter Verification. (If you are going to
perform Smart Meter Verification using the display, only the transmitter and enhanced
core processor versions apply.)
Table 11-1:
Item
Transmitter
Enhanced core processor
ProLink II
ProLink III
Field Communicator
If your devices or tools do not meet the minimum version requirements for Smart Meter
Verification, you still may have access to an older version of meter verification if the option
was ordered for your transmitter. See
between the older meter verification and the newer Smart Meter Verification.
Table 11-2:
Feature
Measurement inter-
ruption
Result storage
Results reporting
Test start methods
11.2.2

Smart Meter Verification test preparation

Although it is not necessary to match factory conditions or change the transmitter's
configuration during a Smart Meter Verification test, the test will run more smoothly when
conditions are stable.
(1) Detailed test analyses such as comparison graphs are not available on the local display.
174
for the minimum version of the transmitter, enhanced core processor, and
Minimum version to support Smart Meter Verification
Minimum version
6.0
3.6
2.9
1.0
HART device description: device rev 6, DD rev 2
Major differences between meter verification and Smart Meter
Verification
Meter verification (older)
Suspended during the test (3 mi-
nutes)
No results stored in the transmitter
Pass/fail/abort
Manual only
Micro Motion
Table 11-2
for a description of the major differences
Smart Meter Verification (newer)
No interruption necessary
Last 20 results stored in the trans-
mitter
Pass/fail/abort, abort code, com-
parison tables and graphs for stor-
ed test results
Manual, scheduled, event-based
®
Model 2700 Transmitters with Analog Outputs