Emerson Micro Motion 2700 Configuration And Use Manual page 186

Transmitters with analog outputs
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

Measurement support
2.
3.
4.
5.
Postrequisites
View the test results and take any appropriate actions.
Run a Smart Meter Verification test using ProLink III
1.
2.
3.
4.
178
You may need to wait a few seconds while ProLink II synchronizes its database with
the transmitter data.
Review the information presented on the screen, and click Next.
Enter any desired information on the Test Definition screen, and click Next.
All information on this screen is optional.
Choose the desired output behavior.
Option
Description
Outputs Continue
During the test, all outputs will continue to report their assigned proc-
Measuring
ess variable. The test will run for approximately 90 seconds.
Outputs Held at Last
During the test, all outputs will report the last measured value of their
Value
assigned process variable. The test will run for approximately
140 seconds.
Outputs Held at Fault
During the test, all outputs will go to their configured fault action. The
test will run for approximately 140 seconds.
Press Start Meter Verification.
Test progress is displayed on the screen.
Choose Device Tools > Diagnostics > Meter Verification > Run Test.
You may need to wait a few seconds while ProLink II synchronizes its database with
the transmitter data.
Enter any desired information on the Test Definition screen, and click Next.
All information on this screen is optional.
Choose the desired output behavior.
Option
Description
Continue Measur-
During the test, all outputs will continue to report their assigned process
ing
variable. The test will run for approximately 90 seconds.
Held at Last Value During the test, all outputs will report the last measured value of their as-
signed process variable. The test will run for approximately 140 seconds.
Held at Fault
During the test, all outputs will go to their configured fault action. The
test will run for approximately 140 seconds.
Press Start.
Test progress is displayed on the screen.
®
Micro Motion
Model 2700 Transmitters with Analog Outputs

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents