Emerson Micro Motion 2700 Configuration And Use Manual page 227

Transmitters with analog outputs
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

Table 12-2:
Status alarms and recommended actions
Alarm code
Description
A117
Density Overrange (Petrole-
um)
A118
Discrete Output 1 Fixed
A120
Curve Fit Failure (Concentra-
tion)
A121
Extrapolation Alarm (Concen-
tration)
A131
Meter Verification in Progress:
Outputs to Last Measured Val-
ue
A132
Sensor Simulation Active
A133
EEPROM Error (Display)
A141
DDC trigger(s) have comple-
ted
N/A
Density FD Calibration in Pro-
gress
N/A
Density D1 Calibration in Pro-
gress
N/A
Density D2 Calibration in Pro-
gress
N/A
Density D3 Calibration in Pro-
gress
N/A
Density D4 Calibration in Pro-
gress
N/A
Zero Calibration in Progress
N/A
Reverse Flow
Configuration and Use Manual
(continued)
Recommended actions
1. Check your process conditions against the values reported
by the flowmeter.
2. Verify the configuration of the petroleum measurement ta-
ble type and density.
The discrete output has been configured to send a constant val-
ue.
Verify the configuration of the concentration measurement ap-
plication.
If the product in the sensor has temperature or density proper-
ties that are outside the parameters established by the concen-
tration measurement curve, this is a symptom and no action is
required. This alarm should clear when the process fluid returns
to normal.
If the problem persists, check the configuration of the concen-
tration measurement application.
Meter verification in progress, with outputs set to Last Measured
Value.
Simulation mode is enabled.
Change the display module. If the problem persists, contact
Micro Motion.
No action required.
No action required.
No action required.
No action required.
No action required.
No action required.
No action required.
No action required.
Troubleshooting
219