For D3 density calibration, the D3 fluid must meet the following requirements:
-
Minimum density of 0.6 g/cm
-
Minimum difference of 0.1 g/cm
density of water. The density of the D3 fluid may be either greater or less than
the density of water.
For D4 density calibration, the D4 fluid must meet the following requirements:
-
Minimum density of 0.6 g/cm
-
Minimum difference of 0.1 g/cm
density of the D3 fluid. The density of the D4 fluid must be greater than the
density of the D3 fluid.
-
Minimum difference of 0.1 g/cm
density of water. The density of the D4 fluid may be either greater or less than
the density of water.
Before performing the calibration, record your current calibration parameters. If the
calibration fails, restore the known values.
11-12.
3
3
between the density of the D3 fluid and the
3
3
between the density of the D4 fluid and the
3
between the density of the D4 fluid and the
®