Chapter 10 Operate The Transmitter With The Weights & Measures Application; Operate The Transmitter When The Weights & Measures Application Is Installed - Emerson Micro Motion 2700 Configuration And Use Manual

Transmitters with analog outputs
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

10
Operate the transmitter with the
Weights & Measures application
Topics covered in this chapter:
Information in this chapter is applicable only if your transmitter was ordered with the
Weights & Measures application.
10.1
Operate the transmitter when the Weights &
Measures application is installed
When the Weights & Measures application is installed, many of the basic functions of the
transmitter are modified.
Configuration and Use Manual
Operate the transmitter when the Weights & Measures application is installed
Switch between secured and unsecured mode
Clear Status Alarm A027: Security Breach
Replacing the core processor in a Weights & Measures installation
To make a connection:
-
If the physical seal is not installed, you can use any supported connection type.
-
If the physical seal is installed, you cannot make a service port connection. You
can use any other supported connection type.
To read or obtain process data:
-
For an approved measurement, you must use an approved method.
-
If the measurement does not have to be approved, you can use any available
method.
Important
The behavior of transmitter outputs and process variable values is modified by the Weights &
Measures application. Ensure that you know the effects of this application on outputs and
process variable values.
To reset totalizers:
-
If the transmitter is unsecured, you can reset totalizers. If you reset one totalizer,
all other totalizers are reset automatically.
-
If the transmitter is secured, you may or may not be able to reset totalizers,
depending on configuration. If totalizer reset is enabled, you cannot reset
totalizers unless the flow rate is 0. To reset totalizers from the display, you may
need to enter the alarm menu password, whether or not you have enabled it. The
default value for Alarm Menu Password is 1234.
Operate the transmitter with the Weights & Measures application
163