Measurement support
Figure 11-10:
D3 or D3 and D4 density calibration using ProLink II
Close shutoff valve
downstream from sensor
11.6.2
Perform a D3 or D3 and D4 density calibration using
ProLink III
Prerequisites
•
•
•
200
D3 calibration
Fill sensor with D3 fluid
ProLink Menu >
Calibration >
Density cal – Point 3
Enter density of D3 fluid
Do Cal
Calibration in Progress
light turns red
Calibration in Progress
light turns green
Close
During density calibration, the sensor must be completely filled with the calibration
fluid, and flow through the sensor must be at the lowest rate allowed by your
application. This is usually accomplished by closing the shutoff valve downstream
from the sensor, then filling the sensor with the appropriate fluid.
For D3 density calibration, the D3 fluid must meet the following requirements:
-
Minimum density of 0.6 g/cm
-
Minimum difference of 0.1 g/cm
density of water. The density of the D3 fluid may be either greater or less than
the density of water.
For D4 density calibration, the D4 fluid must meet the following requirements:
-
Minimum density of 0.6 g/cm
D4 calibration
Done
3
3
between the density of the D3 fluid and the
3
®
Micro Motion
Model 2700 Transmitters with Analog Outputs
Fill sensor with D4 fluid
ProLink Menu >
Calibration >
Density cal – Point 4
Enter density of D4 fluid
Do Cal
Calibration in Progress
light turns red
Calibration in Progress
light turns green
Close
Done