Emerson Micro Motion 2700 Configuration And Use Manual page 96

Transmitters with analog outputs
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

Configure device options and preferences
Effects of Update Rate = Special
Incompatible features and functions
Special mode is not compatible with the following features and functions:
If required, you can switch to Normal mode, perform the desired procedures, and then
return to Special mode.
Process variable updates
Some process variables are not updated when Special mode is enabled.
Table 5-1:
Special mode and process variable updates
Always polled and updated
Mass flow
Volume flow
Gas standard volume flow
Density
Temperature
Drive gain
LPO amplitude
Status [contains Event 1 and Event
2 (basic events)]
Raw tube frequency
Mass total
Volume total
Gas standard volume total
Temperature-corrected volume
total
Temperature-corrected density
Temperature-corrected volume
flow
Batch-weighted average tempera-
ture
Batch-weighted average density
88
Enhanced events. Use basic events instead.
All calibration procedures.
Zero verification.
Restoring the factory zero or the prior zero.
Updated only when the petroleum
measurement application is disa-
bled
RPO amplitude
Board temperature
Core input voltage
Mass inventory
Volume inventory
Gas standard volume inventory
Never updated
All other process variables and calibra-
tion data. They retain the values held
at the time you enabled Special mode.
®
Micro Motion
Model 2700 Transmitters with Analog Outputs