switch from stance to swing. Because DSC is always monitoring knee function, multi-directional movement and
walking backward are also possible without risk of stance resistance releasing.
Adaptive swing phase control
Instantaneous adaptation to varied walking cadences and to changes of the pendular mass (e.g. varying footgear)
ensures the knee always achieves the swing flexion target angle within (+/-) one degree. The swing phase exten­
sion and flexion resistance experienced by the user are auto-adaptive.
The flexed and partially loaded knee will also disable the stance phase on slopes and ramps to allow for greater
knee flexion and more ground clearance in the swing phase.

10 MyModes

In addition to basic mode (mode 1), up to 5 MyModes can be activated and configured with the adjustment soft­
ware. They can be called up by the patient using the Cockpit app. Only the first 3 MyModes can be selected using
movement patterns. Switching by using movement patterns has to be activated in the adjustment software.
These modes are intended for specific motion patterns or postures (e.g. inline skating, running (jogging) ...).
Default settings for these motion patterns and postures can be called up and individually adapted using the adjust­
ment software.
Settings can also be adjusted by the patient using the Cockpit app (see page 42).
Only when "DUAL" marking is not present (see page 12)
For information on switching MyModes with the remote control (optional accessory), see the remote control
instructions for use.

10.1 Running function as configured MyMode

For running over longer periods of time, the "Running" MyMode which can be activated
using the Cockpit app or a movement pattern can be configured using the adjustment soft­
ware.
In this mode, every step will be performed as a running step with larger swing phase angle
and no preflexion at heel strike (PreFlex) (see page 44).
INFORMATION
The running function will work with specialised running feet such as the 1E95 Challenger as well as with prosthet­
ic feet with axial compression such as the 1C61 Triton Vertical Shock. For details on assembly and alignment
please refer to the instructions for use of the foot.
Feet without axial compression are generally not suited for running.

10.2 Switching MyModes with the cockpit app

INFORMATION
Bluetooth on the prosthesis must be turned on in order to use the Cockpit app.
If Bluetooth is switched off, it can be turned on by turning the prosthesis upside-down (function only available in
basic mode) or by connecting/disconnecting the battery charger. Bluetooth is then turned on for approx. 2
minutes. During this time, the app must be started and used to establish a connection. If required, Bluetooth on
the prosthesis can be switched on permanently afterwards (see page 43).