Equipment Classification

Per EN/IEC 60601-1, EN/IEC 60601-2-54, this device is classified as
following:
Table 5: Equipment classification
Class I equipment
Type B equipment
Water ingress
Cleaning
Disinfection
Flammable anesthetics
Operation
Related Links
Cleaning and Disinfecting
3255F EN 20210312 1125
Equipment in which protection against electric
shock does not rely on basic insulation only, but in-
cludes a fixed connection to mains power with pro-
tective earth conductor.
A Type B piece of equipment is one that provides a
particular degree of protection against electric
shock particularly regarding allowable leakage cur-
rent and reliability of the protective earth protec-
tion.
IP10
This device does not have protection against in-
gress of water.
See section on cleaning and disinfecting.
See section on cleaning and disinfecting.
This device is not suitable for use in the presence of
a flammable anesthetic mixture with air, or in
presence of a flammable anesthetic mixture with
oxygen or nitrous oxide.
Continuous operation.