20 | Full Leg Full Spine DR Retrofit System | Introduction to the Full Leg Full Spine application

Equipment Classification

Per EN/IEC60601-1, this device is classified as following:
Table 2: Equipment classification
Class I Equip-
ment
Type B Equip-
ment
Water Ingress
Cleaning
Disinfection
Flammable An-
esthetics
Continuous Op-
eration
Expected serv-
ice life
0326A EN 20211014 1141
Equipment in which protection against electric shock does
not rely on basic insulation only, but includes a power sup-
ply cord with protective earth conductor. For earth relia-
bility always plug the main power cord into an earthed
mains power outlet.
A Type B applied part is one that provides a particular de-
gree of protection against electric shock particularly re-
garding allowable leakage current and reliability of the
protective earth protection.
This device does not have protection against ingress of wa-
ter.
See section on maintenance and cleaning.
See section on maintenance and cleaning.
This device is not suitable for use in the presence of a
flammable anesthetic mixture with air, or in presence of a
flammable anesthetic mixture with oxygen or nitrous ox-
ide.
The unit is suitable to run continuously.
Minimum 7 years.