Mx840B-R Universal Module - HBM SomatXR CX23-R User Manual

Somatxr series ultra-rugged daq
Hide thumbs Also See for SomatXR CX23-R:
Table of Contents

Advertisement

5   SomatXR hardware
  Transducer
Strain gage, full bridge
six-wire configuration
Strain gage, half bridge
five-wire configuration
Strain gage, quarter bridge
three- or four-wire configuration
Inductive full bridge
Inductive half bridge
LVDT
(linear variable differential transformer)
84
  MX1615B Bridge Module (Strain Gage Amplifier): Up to 16 individually
configurable inputs, including strain gage-based transducers, standardized
voltage and resistance or resistance-based measurements. This module is
functionally equivalent to the MX1601B-R in laboratory environments with the
following exceptions – 15V input mode instead of 60V input mode, one positive
swing (upscale polarity) shunt resistor only, shunt scaling is not supported.
  MX471B CAN Module: The module has 4 CAN bus nodes that can be
configured for receiving messages.
  MX878B Analog Output Module: Up to 8 analog output channels with +/-10V
range that can be sourced from any MX module channel that is connected via
FireWire. CX23-R/eDAQXR support is limited to an input follower mode where
each analog output channel follows the defined input data source using the
following scaling. The user defined Range max parameter value for the source MX
channel maps to +10V analog output voltage; the user defined Range min
parameter value maps to -10V analog output voltage. The nominal data skew
between the analog output and the source input to the associated MX module
channel is nominally 3 milliseconds when no digital filters are used.
  MX840B Universal Module: Up to eight (8) universal inputs for connecting more
than 16 transducer technologies. This module is functionally equivalent to the
MX840B-R in laboratory environments with the following exception – one positive
swing (upscale polarity) shunt resistor only.
  MX460B Frequency / Counter Module: Up to four (4) digital inputs to measure
digital pulses up to 1 MHz (speed, torque, angle position, displacement, PWM in
general). This module is functionally equivalent to the MX460B-R in laboratory
environments. CX23-R/eDAQXR support is limited in the following aspects. Math
functions are not supported. Interactive zeroing of the "crank shaft" sensors is not
supported.
5.3.1   MX840B-R Universal Module
Up to eight (8) universal inputs can be connected to the MX840B-R module
compatible with more than 16 transducer technologies.
a4275-8.0 en    HBM: public
MX840B-R
Wiring diagram
*
    ●   
3-wire only
CX23-R and EX23-R
99
99
99
102
102
102

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Somatxr ex23-r

Table of Contents