4   Setting up and running a test
64
 
After true - false edge: Start on the sample after the input channel transitions
o
from TRUE to FALSE.
Trigger delay setup
  Period (secs): Enter the number of seconds after the trigger start to set the output
channel state.
  Conditional mode: Set the output channel state based on the behavior of the
input channel during the delay period.
 
Unconditional: Set the output to TRUE regardless of the input channel.
o
Trigger sustain setup
  Period (secs): Enter the number of seconds to sustain the output channel state.
At the end of the sustain period, the output value is set to FALSE and the system
waits for the next trigger start condition.
  Conditional mode: Set the output channel state based on the behavior of the
input channel during the sustain period.
 
Unconditional: Set the output to TRUE for the duration of the sustain period
o
regardless of the input channel.
4.4.20   Track
The Track computed channel generates an output channel that tracks the minimum
value, maximum value or range of the input channel. A logical trigger channel can be
used to reset the tracker.
  Mode: Set the tracker mode to Min track, Max track or Range track.
Trigger reset options
  Enable triggered reset: If selected, the tracker resets when the specified
Channel meets the Reset mode condition. For Min track and Max track modes,
resetting the tracker sets the output to the current sample value. For the Range
track mode, resetting the tracker sets the output to zero.
  Channel: Set the trigger channel used for the Enable triggered reset option.
  Reset mode: If the Enable triggered reset option is selected, set the condition for
the trigger channel to reset the tracker.
 
When true: Reset when the trigger channel is TRUE.
o
 
On false - true edge: Reset when the trigger channel transitions from FALSE
o
to TRUE.
 
After true - false edge: Reset on the sample after the trigger channel
o
transitions from TRUE to FALSE.
4.4.21   Triggered latch
The Triggered latch computed channel generates an output channel that latches the
previous input channel data value when triggered.
Trigger
  Channel: Set the trigger channel used to latch input channel data. By default, the
channel latches the input data value when the trigger channel transitions from
FALSE to TRUE and holds this data value until the trigger channel transitions
a4275-8.0 en    HBM: public