4   Setting up and running a test
40
  Filter frequency: Set the frequency of the selected filter. The options vary based
on the filter type.
  Custom column information: Enter the desired information. The system treats
all input as free formatted text. This means it will sort alphabetically, but not
alphanumerically (e.g., input values of "1", "2", "11", will be sorted descending as
"1", "11", "2"). Use leading zeros to realize alphanumeric sorting.
  Range min / Range max: Enter the desired full scale range min and max values
in physical (engineering) units. For all channels, they are used as the default min
and max bounds for histogram DataModes.
  Alarm min / Alarm max: Enter the desired values for min and/or max alarms as
a physical units value or as a % of Range min/max value.
  Warning min / Warning max: Enter the desired values for min and/or max
warnings as a physical units value or as a % of Range min/max value.
  Chart type: Set the default chart type for the channel. Not in edit dialog.
  Decimals: Select the default decimal places for channel displays. Use the "Auto"
option to apply the settings defined in the User Preferences. See
preferences
> General > Decimal point handling for more information.
4.3.2   Input channel scaling parameters
Before running a test, many input sensors must be scaled. Define sensor scaling
parameters in
Input
To configure a channel for sensor scaling, first set the Scaling mode in the channel
editor dialog or channel grid.
Depending on the mode selected, different scaling parameters must be specified to
define the scaling.
Sensor scaling mode
Defined slope intercept
Defined zero span
Defined two point
Experimental two point
Internal shunt resistor
Strain gage
Scaling parameters
  Scaling mode: Set the channel scaling mode.
a4275-8.0 en    HBM: public
channels.
Scaling parameters
Slope
Intercept
Electrical zero
Electrical span
Physical span
Electrical 1
Physical 1
Electrical 2
Physical 2
See
Experimental two point scaling
See
Internal shunt resistor scaling
Strain gage factor
Bridge factor
User
CX23-R and EX23-R