3   Configuring the system
CX23-R and EX23-R
3.1   Configuring network settings
The data processor communicates through standard 10/100/1000 Base-T Ethernet
using IPv4 or IPv6. The unit hosts its own secure web interface accessed through
configurable network settings. The default IP address is 192.168.100.101.
Defining a network configuration
The data processor uses network configurations to determine how to boot up and
communicate with the browser. Many network configurations can be defined, but
only the active configuration is considered at boot time.
View and manage network configurations on the Networks page under System
Configuration.
To define a new network configuration, click New. Select Static or Dynamic as
desired.
Select Static and enter the appropriate parameters:
  Network: Enter a full description of the network. This is the text that displays in
the networks list.
  IP address assignment: Set the type of IP address to either Static or Dynamic.
For static assignment, enter correctly formatted IPV4 address, subnet mask,
default gateway and broadcast values. Some fields are automatically populated
based on the entered values, but they may be edited as necessary.
  DNS server assignment: For static IP assignment, manually format the
Preferred and Secondary DNS servers.
Select Dynamic - Obtain IP address configurations automatically (DHCP)
(Dynamic Host Configuration Protocol), to get information such as an IP address,
subnet mask and default gateway. If on a network with a DHCP server, use the
HBM Device Manager (on the included USB drive) or contact your IT administrator to
find out the IP address. If presented with a certificate warning, choose to proceed.
Click Save to save the new configuration. To make the new configuration active,
click the Make active button. The unit reboots with the new settings.
Default network configuration and restoring default settings
The unit comes factory set with the IP address 192.168.100.101. The default active
network configuration is named Default Auto Set and is configured for dynamic IP
address assignment (DHCP). the default network cannot be deleted.
If it becomes necessary to restore the default settings, there is a reset procedure that
sets the IP address to 192.168.100.101 and makes the Default Auto Set the active
network. No other network configurations are affected. To perform the reset:
 1. Save all changes and ensure there are no other users on the system.
 2. Switch off the unit by pressing the power switch and wait for the unit to
completely power down.
 3. Press and hold the power switch.
a4275-8.0 en    HBM: public