Connector
pin side of cable
Transducer
Voltage,
60 V, 10 V, 100 mV
Current-fed piezoelectric transducer
(IEPE, ICP®)
CX23-R and EX23-R
Pin
Connection
1
Excitation (-)
jumper to pin 11
2
Excitation (+)
3
Voltage input 60 V (+)
4
Signal ground
5
--
6
--
7
Measurement signal (+)
8
Measurement signal (-)
9
--
10
--
11
Sense (-)
jumper to pin 1
12
Sense (+)
13
TEDS (-)
14
TEDS (+)
Shield
Shield
NOTE
Pin assignment is different for
measurements
information, refer to the wiring diagrams.
NOTE
Connection between pins 1 and 11 is necessary for all MX1615B-R transducers.
Note that the sensor connector must have a connection between pins 4 and 5 for
compatibility with other MX modules.
5.3.3 MX1601B-R Standard Module (Universal Amplifier)
Up to 16 configurable inputs for DC voltage sources (60 V, 10 V, 100 mV), DC
current sources (20 mA) or current-fed piezoelectric transducers (IEPE, ICP®) can
be connected to the MX1601B-R module.
a4275-8.0 en HBM: public
Strain gage
and
Resistance-based measurements
MX1601B-R
5 SomatXR hardware
Wire color
(1-KAB183 or 1-KAB184)
Black
Blue
White/Black
Red/Black
Pink/Black
Yellow/Black
White
Red
Brown
Yellow
Grey
Green
Grey/Black
Green/Black
--
transducers,
Resistance-based
inputs. For more
Wiring diagram
●
†
●
108
109
89