Data Storing - EDAN M3 User Manual

Vital signs monitor
Hide thumbs Also See for M3:
Table of Contents

Advertisement

M3 Vital Signs Monitor User Manual
scanner. If it is set to OFF, a message box indicating "Confirm to update patient,
yes?"will pop up when scanning a barcode. Click on YES to automatically update patient
information; click on NO to quit automatical update.
ADU, PED, NEO:
Select a digit from 0~9 to indicate the patient type. For example, if ADU is set to 9, PAT
TYPE is set to 1, and the first digit in the barcode is 9, the patient type could be updated to
ADU.
NOTE:
1
The set value of START/END in ID as well as the set value of PAT TYPE must not
exceed the length of the barcode.
2
If START/END is set to 0, the patient ID will not be updated by using barcode
scanner.
3
If PAT TYPE is set to 0, the patient type will not be updated by using barcode
scanner.
4
Barcodes containing characters other than digits or containing space will be
considered invalid and cannot be identified. If any invalid character is detected, a
message box indicating "Special signs are in code bar!" will pop up.
5
Connect the barcode scanner to the monitor and wait 10 seconds before starting the
scanner.
6
Refer to the accompanying operator's manual of the scanner for more information
about its usage.
EXIT: Exit the menu.
Factory Maintain
Factory maintenance function is only available for the service engineers of EDAN or
representative authorized by EDAN.

4.12 Data Storing

The monitor can support the USB storage for the Data Store function. Enter the menu by
SYSTEM MENU > DATA STORE to set the data storing function. You can set the storing
interval, browse data, search data, and delete all the data or single item data.
- 42 -
System Menu