M3 Vital Signs Monitor User Manual
The sound pressure range for audible alarm signals is from 45 dB to 85 dB.
Do not rely exclusively on the audible alarm system for patient monitoring. Adjustment of
alarm volume to a low level or off during patient monitoring may result in patient danger.
Remember that the most reliable method of patient monitoring combines close personal
surveillance with correct operation of monitoring equipment.
NOTE:
1
The monitor does not have alarm condition delay or alarm signal generation delay.
2
When alarms of different levels occur at the same time, the monitor prompts one of
the highest levels.
3
If the monitor is powered off and then turned on, the alarm setup can resume to the
setup which is set before the power-off.

5.1.3 Alarm Setup

Select the ALARM SETUP in the SYSTEM MENU to open the submenu as shown below.
Alarm setup of each parameter
You can turn ON or OFF the alarm for each parameter, and set the upper and lower alarm limit
for each parameter by ALM HI or ALM LO.
In the ALARM SETUP menu, set the alarm limit for each parameter: SYS, DIA, MAP, SpO
WARNING
Figure 5-1 Alarm Setup
- 49 -
Alarm
,
2