A1.2.9 Temp (Optional); A1.2.10 Wirless Network - EDAN M3 User Manual

Vital signs monitor
Hide thumbs Also See for M3:
Table of Contents

Advertisement

M3 Vital Signs Monitor User Manual

A1.2.9 TEMP (Optional)

T2 Module:
Measuring range
Working temperature
Sensor type
Alarm range
Resolution
Accuracy
Response time
Update time
TH Module:
Measuring range
Working temperature
Alarm range
Resolution
Accuracy
Response time

A1.2.10 Wirless Network

Compliant with Standard and
Directive
Frequency Range
25°C ~ 45°C
10°C ~ 40°C
Oral /axillary /rectal
35.5°C ~ 42°C
0.1°C
±0.1°C (25°C ~ 45°C)
< 60s
1s ~ 2s
34°C ~ 42.2°C
10°C ~ 40°C
35.5°C ~ 42°C
0.1°C
±0.2°C (35.5°C ~ 42°C)
±0.3°C (out of the limits)
1s
IEEE802.11b/g, R&TTE Directive (99/5/EEC)
2.412 GHz ~2.462 GHz (America)
2.412 GHz ~2.484 GHz (Japan)
2.412 GHz ~2.472 GHz (ETSI)
- 102 -