Lead Input/Usb Socket; Control Key; Data Storage - EDAN SE-2003 User Manual

Holter system recorder
Table of Contents

Advertisement

SE-2003&SE-2012 Holter System Recorder User Manual

2.2 Lead Input/USB Socket

SE-2003/SE-2012 is featured by a 26-pin lead input/USB multi-functional connector socket. It
serves as lead input socket when collecting records. After recording finishes, insert USB cable
into the socket and data could be read into analysis system through USB 2.0 high speed transfer.
By using the same socket, possibility of connection of lead and USB with recorder at the same
time is prevented and potential safety hazards are effectively avoided.
NOTE: Pay attention to the plug direction. Please insert the side with marker toward
recorder with appropriate force.
CAUTION
Do not sway plug during use. Plug may fall off and that will cause record failure and even
damages to socket.

2.3 Control Key

SE-2003/SE-2012 keyboard is composed of confirmation key, left-up key and right-down key.
Confirmation key is used to confirm choice operations and change cursor status in menu
operation mode.
Confirmation key serves as patient events key in monitoring process. If patients feel
uncomfortable or want to record time of symbolic meaning (start to do exercise, begin to sleep,
etc), press the key and the recorder will record the time.
Left-up key is used to move the cursor to left or move up in menu operation mode.
Right-down key is used to move the cursor to right or move down in menu operation mode.
For detailed functions of different buttons, please refer to parameter setting section.

2.4 Data Storage

SE-2003/SE-2012 stores ECG data in Secure Digital card (SD card) which will be analyzed by
Holter analysis software after finishing recording.