A2.3 Electromagnetic Immunity - For Equipment And Systems That Are Not Life-Supporting - EDAN M3 User Manual

Vital signs monitor
Hide thumbs Also See for M3:
Table of Contents

Advertisement

M3 Vital Signs Monitor User Manual
A2.3 Electromagnetic Immunity - For EQUIPMENT and
SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The M3 is intended for use in the electromagnetic environment specified below. The customer
or the user of M3 should assure that it is used in such an environment.
Immunity
IEC/EN 60601 test
test
Conducted
3 Vrms
RF IEC/EN
150 kHz to 80 MHz
61000-4-6
Radiated RF
3 V/m
IEC/EN
80 MHz to 2.5 GHz
61000-4-3
Compliance
level
level
3Vrms
3 V/m
Electromagnetic environment
Portable and mobile RF communications
equipment should be used no closer to
any part of the EUS T Ultrasound
Scanner, including cables, than the
recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
3
5 .
=
d
V
1
3
5 .
=
d
E
1
7
=
d
P
E
1
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
less than the compliance level in each
frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
- 106 -
EMC Information
-guidance
P
P
80 MHz to 800 MHz
800 MHz to 2.5 GHz
a
should be
b