Advertisement

M3 Vital Signs Monitor User Manual
6.
If the predict measuring is successfully finished, the monitor enters MONITOR mode after
30s; otherwise the monitor enter MONITOR mode immediately after the predict measuring.
The monitoring state lasts for 10 min, and then the monitor enters waiting state.
displays in the TEMP parameter area on interface. Put the sensor back into the sensor
bracket.
7.
If necessary, repeat the measurement according to the procedure above.
NOTE:
After one measurement, the user should put the sensor back to the sensor bracket and
then take it out for starting a new measurement.
The monitor's state can change from the PREDICT mode into the MONITOR mode, but it can
not change from the MONITOR mode into the PREDICT mode.

10.1.3 TEMP Setup Menu

10.1.3.1 TEMP Setup
Click on the TEMP SETUP in the SYSTEM MENU to display the following figure:
MEASURE MODE: Set the measuring mode to PREDICT or MONITOR.
MEASURE POS: Set the measuring position to ORAL, AXILLARY or RECTA. The
axillary sensor can be used for measuring oral/axillary temperature, while the rectal sensor
for measuring rectal temperature.
10.1.3.2 TEMP Unit Setup
If you want to change the TEMP unit, please select SYSTEM MENU >MAINTAIN >ENTER
MAINTAIN PASSWORD >USER MAINTAIN and you will see the item TEMP UNIT on the
menu as follows:
Figure10-2 TEMP Setup
- 79 -
TEMP Monitoring (Optional)
— — —