System Menu
User Maintain
Input the user password 9 9 8 1 in the USER KEY box and press OK, USER MAINTAIN menu
will pop up, in which you can set following items.
Figure 4-21 User Maintain
BED No.: Set the bedside number to a value from 1 to 64.
LANGUAGE: Set the displayed language.
NOTE:
You should restart the monitor after changing the displayed language to make the
operation effective.
NURSE CALL: Turn on or off the nurse call. When the parameter alarm occurs, the monitor
gives 3s nurse call alarm prompt; if the audible alarm or the audio system is off, the monitor can
also give the nurse call alarm in abnormal condition.
The relay contact between pin7 and pin8 of RJ45 is normally open. But it is closed when an
alarm is audible.
SERVER IP: The default server IP is 202.114.4.119. It can be changed by the user according to
the IP of PC installed with MFM-CMS of EDAN.
SERVER PORT: Set server port.
PRES UNIT: Set the pressure unit to mmHg or kPa.
TEMP UNIT: Set the temperature unit to °C or℉.
HOT KEY: Set the hot key to PATIENT or PRINT.
- 37 -