EDAN M3 User Manual page 93

Vital signs monitor
Hide thumbs Also See for M3:
Table of Contents

Advertisement

M3 Vital Signs Monitor User Manual
2
For children over two-year old and adults: pull the ear straight up and back as shown
below:
5.
Press the "Scan" button for one second until you hear a long beep sound which signals the
end of the measurement, and rusults will be shown on the display of the monitor.
6.
Before starting another measurement, wait until all icons stop flashing and two beep sounds
are heard.
Replace the probe cover after each use to ensure an accurate reading and avoid cross
contamination.
NOTE:
1
The thermometer will automatically shut down after one-minute pending to extend
battery life.
2
The device must stay in stable ambient (room) temperature for 30 minutes before
operation.
3
Before the measurement, please stay in a stable environmrnt for five minutes and
avoid exercise or bath for 30 minutes.
4
It is recommended that you measure the same ear for three times. If the three
measurements are different, select the highiest temperature.
5
Remember to compare the measurement result to the regular temperature of the
patient.
6
There is no gender and age limitation for using infrared ear thermometer.
7
The data saved in the thermometer is the last measurement data before the
thermometer is powered off.
8
Clinical repeatability: 0.18
old).
WARNING
(<1 year old)
; 0.12
- 87 -
TEMP Monitoring (Optional)
(1~5 years old); 0.1
0
(>5 years