M3 Vital Signs Monitor User Manual
MOTION
noise is too large or pulse
rate is not regular.
OVER
Pressure has exceeded the
PRESSURE
specified upper safety limit.
SIGNAL
Excessive motion.
SATURATED
PNEUMATIC
During pneumatic test, leak
LEAK
is detected.
CUFF
TYPE
Cuff type does not comply
ERR
with the patient type.
NIBP
TIME
Measuring time has exceeded
OUT
120s (adult) or 90s (neonatal).
Prompt message: (display in the prompt area below NIBP value):
Message
Manual measuring...
Auto measuring...
Measurement over
Calibrating...
Calibration over
Pneum testing...
Pneum test over
Resetting...
Reset failed
Low
Low
Low
Low
Low
Cause
During manual measuring mode.
During automatic measuring mode.
Measurement over
During calibrating
Calibration over
During pneumatic test
pneumatic test over
NIBP module in resetting
NIBP module reset failed
- 74 -
NIBP Monitoring (Optional)
under
monitoring
motionless.
Measure again, if failure
persists,
stop
measuring function of NIBP
module
and
biomedical
engineer
manufacturer's service staff.
Stop
the
patient
moving.
Check
and
replace
leaking parts, if required,
notify biomedical engineer or
manufacturer's service staff.
Select appropriate cuff type
Measure again or use other
measuring method.
Alarm Level
No alarm
is
using
notify
or
from
the