M3 Vital Signs Monitor User Manual
Size of reusable cuff for neonate/pediatric/adult:
Patient Type
Infant
Child
Adult
Large Adult
Thigh
Size of disposable cuff for neonate/children/adult:
Size No.
1
2
3
4
The lifespan of cuff is: 480mmHg/20000 times; 300mmHg/50000 times.
Make sure that the cuff edge falls within the range of mark < >. If it does not, use a
larger or smaller cuff that fits better.
3. Connect the cuff to the air hose. The limb chosen for taking the measurement should be
placed at the same level as the patient's heart. If this is not possible you should apply the
following corrections to the measured values:
If the cuff is placed higher than the heart level, add 0.75 mmHg (0.10 kPa) for each inch
of difference.
If it is placed lower than the heart level, deduct 0.75 mmHg (0.10 kPa) for each inch of
difference.
4. Check whether the patient mode is appropriately selected. Access PATIENT SETUP menu
from SYSTEM MENU and pick PAT TYPE item and select the required patient type.
5. Select a measurement mode in the NIBP SETUP menu. Pick the INTERVAL item for
MANUAL or set the interval for auto measurement; or select the CONTINUAL mode.
6. Press the NIBP START/STOP on the front panel to start a measurement. You can also stop
this measurement by this button.
Limb perimeter
10 cm ~19 cm
18 cm ~ 26 cm
25 cm ~ 35 cm
33 cm ~ 47 cm
46 cm ~ 66 cm
Limb perimeter
3.1 cm ~ 5.7 cm
4.3 cm ~ 8.0 cm
5.8 cm ~ 10.9 cm
7.1 cm ~ 13.1 cm
- 67 -
Cuff width
8 cm
10.6 cm
14 cm
1.5 m or 3 m
17 cm
21 cm
Cuff width
2.5 cm
3.2 cm
1.5 m or 3 m
4.3 cm
5.1 cm
NIBP Monitoring (Optional)