EDAN M3 User Manual page 89

Vital signs monitor
Hide thumbs Also See for M3:
Table of Contents

Advertisement

M3 Vital Signs Monitor User Manual
Probe data error
Probe heater error
Probe temp too
high
Temp
SENSOR
OFF
Prompt:
Message
Warm-up over
Measure over
Measure time out
Offline:
NTC
resistance >R 0 ℃;
Short:NTC resistance
<R+100
.
Single failure
The
original
temperature of sensor
>+33 ℃ & ≤+40
After
the
sensor
temperature reaches
Predict
value,
descends to the value
lower than Predict
value.
Cause
The monitor prompts it after taking
the sensor out of the bracket and
warm-up is over.
After the Predict measuring is
over, the data and message display
on the interface.
No measuring result after the
module entering Predict state for
30s.
Med
Med
Low
.
Med
it
Put the sensor to the measuring
position and start measuring.
Enter monitoring state after the
Predict state. After monitoring for
10 min, it returns to waiting state.
After monitoring for 10 min, it
returns to waiting state.
- 83 -
TEMP Monitoring (Optional)
Put the sensor into the sensor
bracket,
take
it
measure again. If the problem
persists, stop using measuring
function of TEMP module,
notify biomedical engineer or
Manufacturer's service staff.
Put the sensor into the sensor
bracket,
take
it
measure again. If the problem
persists, stop using measuring
function of TEMP module,
notify biomedical engineer or
Manufacturer's service staff.
Put the sensor into the sensor
bracket, measure again after
the sensor temperature reaches
normal value.
Reconnect the sensor and make
sure that the cable is properly
connected.
What to do
out
and
out
and