10.1.5 Care and Cleaning

Before cleaning the monitor or the probe, make sure that the equipment is switched off
and disconnected from the power line.
Reusable TEMP Probes
The TEMP probe should not be heated above +100 ℃ (+212 ℉). It should only be briefly
1
exposed to temperatures between +80 ℃ ~ +100 ℃ (+176℉ ~ +212℉).
2
The probe must not be sterilized in steam.
3
Only detergents containing no alcohol can be used for disinfection.
4
All the sensors should be used with a protective rubber cover.
5
To clean the probe, hold the tip with one hand and rub the probe down from the connector
with the other hand using a moist lint-free cloth.
NOTE:
1
Wash the probe with clean water after disinfecting and sterilizing to remove any
remaining solution. The probe can only be reused after being dried thoroughly.
2
Do not disinfect the probe by means of boiled water.
3
The product has not been disinfected at the factory.
4
Any residue should be removed from the probe before being disinfected and
sterilized, and avoid contacting corrosive solvent. Dipping the cable into alcohol or
alkalescent solvent for a long time may reduce the flexibility of the scarfskin of the
cable. Also, the connector should not be dipped.
5
After monitoring, disinfect the probe according to the instruction described in the user
manual.
6
Cavity temperature probe is suggested to be used only inside the recta. It is
recommended to use the disposable cannula to prevent cross infection.
7
For protecting the environment, the disposable TEMP probe cover must be recycled
or disposed of properly.
8
Do not force the cavity temperature probe against resistance when inserted into
human body. Also it is not recommended to use it in bleeding part or cankerous part
of human body.
WARNING
- 84 -
TEMP Monitoring (Optional)