M3 Vital Signs Monitor User Manual
NIBP MEMORY
You can set this item to ON or OFF. If the item is ON, the monitor will automatically
memorize the initial measurements of the patient when measuring his or her blood pressure.
Then the monitor will inflate the cuff according to the previous memorized measurements.
This function accelerates the measuring of the patient's blood pressure.
ALARM SETUP:
ALARM NUTE: Set the duration of silencing the audible alarm to 60s, 120s or 180s.
ALARM SILENCE
You can set this item to ON or OFF. If the item is ON, you can turn off the audio system by
pressing the SILENCE button on the front panel for more than 2s. In this case, all sounds
including the alarm sound, key sound and sphygmic sound coming from the monitor will be
mute. If the item is OFF, the function mentioned above is unavailable.
BARCODE SETUP:
ID:
Patient ID can maximumly be a three-digit number. On this precondition, you can determine
which digit in the barcode is the starting/ending digit for the paitent ID via configuration of
START and END. Take the following barcode for example. If you set START to 2 and
END to 4, the updated patient ID will begin with the second digit and end with the fourth
digit in the barcode, namely 787.
PAT TYPE:
You can determine which digit in the barcode indicates the patient type. For example, if you
set PAT TYPE to 1, the first digit in the barcode will be identified as an indication of
patient type.
ON/OFF:
If it is set to ON, the patient information is updated automatically by using a barcode
Figure 4-26 Diagram of NIBP Air Leakage Test