EDAN M3 User Manual page 10

Vital signs monitor
Hide thumbs Also See for M3:
Table of Contents

Advertisement

M3 Vital Signs Monitor User Manual
23 When the monitor and electrosurgical device are used together, the user (physician
or nurse) should guarantee the safety of patient.
24 Please disinfect timely to prevent cross infection between patients.
25 This monitor is not a device for treatment purposes.
26 Only NIBP and SpO
defibrillator is applied, keep other accessories away from the patient. Otherwise, it
may result in damaging the monitor or harming the patient.
27 Do not touch the patient, bed or instrument during defibrillation.
28 During monitoring, if the power supply is off and there is no battery for standby, the
monitor will be off, and only the patient information and alarm settings can be saved.
After reconnecting the power supply, the user should turn on the monitor for
monitoring.
1
Federal law (U.S.) restricts this device to sale by or on the order of a physician.
2
Electromagnetic Interference – Ensure the environment in which the monitor is
installed is not subject to any sources of strong electromagnetic interference, such as
radio transmitters, mobile telephones, etc.
3
The monitor is designed for continuous operation and is "ordinary" (i.e. not drip or
splash-proof).
4
Keep the environment clean. Avoid vibration. Keep it far from corrosive medicine,
dust area, high-temperature and humid environment.
5
Do not immerse transducers in liquid. When using solutions, use sterile wipes to
avoid pouring fluids directly on the transducer.
6
Do not sterilize the monitor, recorder or any accessories.
7
The device and accessories are to be disposed of according to local regulations after
their useful lives. Alternatively, they can be returned to the dealer or the manufacturer
for recycling or proper disposal.
8
Remove a battery whose life cycle has expired from the monitor immediately.
9
Avoid liquid splash and excessive temperature. The temperature must be kept
between +5
+55
during transportation and storage.
10 Before use, the equipment, patient cable and sensor should be checked.
Replacement should be taken if there is any evident defectiveness or aging symptom
which may impair the safety or performance.
applied parts of the monitor are defibrillation-proof. When a
2
and
+40
while working. And it should be kept between
WARNING
CAUTION
- 4 -
Intended Use and Safety Guidance
-20
and