NOTE:
The concrete presentation of each alarm prompt is related to the alarm level.
Screen Display
When the measured parameter exceeds its alarm limits and triggers a physiological alarm, the
monitor will give alarm prompt on the screen indicating the occurrence of alarm.
The description will display in Information area, such as "**NS TOO HIGH", and
displays beside the parameter to indicate the low-medium level alarm.
Technical alarm will not prompt * signal.
Alarm Level
High
Medium
Low
Lamp light
The high/medium/low-level alarms are indicated by the system in following different visual
ways:
Alarm Level
High
Medium
Low
Alarm Sound
The high/medium/low-level alarms are indicated by the system in following different audio ways:
Alarm Level
Mode is "beep-beep-beep------beep-beep, beep-beep-beep------beep-beep",
High
which is triggered once every 5 s.
Medium
Mode is "beep-beep-beep", which is triggered once every 20 s.
Low
Mode is "beep-", which is triggered once every 25 s.
1:
displays in Parameter area
2: *** displays beside the parameter (Physiological alarm only)
1:
displays in Parameter area
2: ** displays beside the parameter (Physiological alarm only)
1:
displays in Parameter area
2: * displays beside the parameter (Physiological alarm only)
Alarm indicator flashes in red with high frequency.
Alarm indicator flashes in orange with low frequency.
Alarm indicator lights on in orange.
Visual Prompt
Visual Prompt
Audio Prompt
- 48 -
Alarm

Advertisement