M3 Vital Signs Monitor User Manual
Technical alarms:
Message
SpO
SENSOR OFF
2
SEARCH PULSE
SpO
COMM STOP
2
SpO
LOW
2
PERFUSION
NO SpO
SENSOR
2

8.7 Maintenance and Cleaning

1
Before cleaning the monitor or the sensor, make sure that the equipment is switched
off and disconnected from the power line.
2
Do not subject the sensor to autoclaving.
3
Do not immerse the sensor into any liquid.
4
Do not use any sensor or cable that may be damaged or deteriorated.
Cause
The SpO
sensor may be
2
disconnected
from
patient.
SpO
sensor
may
2
disconnected
from
patient or the monitor.
SpO
module failure or
2
communication failure
The
measured
signals
coming from pulse are too
weak.
The
SpO
sensor
2
disconnected
from
monitor, or the sensor is
not connected well to the
device.
WARNING
Alarm
Level
High/ Low
the
(Configured
by the user)
be
the
Low
High
Low
is
the
Low
- 63 -
SpO
Monitoring (Optional)
2
What to do
Make sure the sensor is
attached to the patient's
finger
or
another
appropriate position.
If no pulse information is
displayed after 30s passes,
check whether the sensor
is connected with the
patient's
finger.
necessary,
attach
sensor to another position
which
might
provide
better signal.
Stop
using
measuring
function of SpO
module;
2
notify biomedical engineer
or manufacturer's service
staff.
Reconnect the sensor, or
choose another measured
position. If the problem
remains,
please
biomedical engineer or
manufacturer's
service
staff.
Reconnect the sensor with
the monitor. Make sure
the
monitor
is
connected with the cable.
If
the
notify
well