Laser On/Off Status

Usage:
Laser ON/OFF extends the life of the laser. It is useful where there
are significant time gaps between symbols. It also provides visual con-
firmation of a read cycle duration and minimizes laser exposure to
people.
Definition:
When enabled, the laser is ON only during the read cycle. When disabled,
the laser operates continuously.
Serial Cmd:
<KClaser on/off
Default:
Disabled
Options:
0 = Disabled
Note: A serial or external trigger must be enabled for Laser ON/OFF to take effect.
Note: Laser ON/OFF is not recommended when New Trigger is enabled, since the
laser will effectively remain on continuously.
MS-710 Scanner User's Manual
Chapter 6 I/0 Parameters
status>
1 = Enabled
6-15