Raman ASTM Signal-to-Noise Test

These tests verify the Raman ASTM signal-to-noise for
• single wavelength (at signal) (EX=350 nm, EM=397 nm) or
• single wavelength (at background) (Ex=350 nm, Em=397 nm, dark value 450
• dual wavelength (EX=350 nm, EM
The specification single wavelength at signal can be measured with the Agilent LabAdvisor.
N O T E
All other have to be set up manually as described for
page 160 with the information from
Table 29
Duration
Report Style (Agilent ChemStation)
Noise Determination
Solvent
Flow rate
Specification (single wavelength at signal)
Specification (single wavelength at
background)
Specification (dual wavelength)
Agilent 1260 FLD User Manual
nm) or
Raman Signal-to-Noise Test Conditions

Raman ASTM Signal-to-Noise Test

=397 nm, EM
A
B
"Dual Wavelength Verification"
Table 31
on page 156 and
approximately 23 minutes
Performance+Noise
5 to 20 minutes
LC grade water, degassed
0.5 - 1 ml/min
>500 (according to settings in
page 156)
>3000 (according to settings in
page 156)
>300 (according to settings in
page 156)
Test Functions
=450 nm)
Table 32
on page 156.
Table 30
on
Table 31
on
Table 32
on
8
on
155