Run Recovery (G1314E) - Agilent Technologies 1200 Infinity Series User Manual

Hide thumbs Also See for 1200 Infinity Series:
Table of Contents

Advertisement

4
Using the Detector
Special Settings of the Detector

Run Recovery (G1314E)

This feature is not supported with ChemStation B.03.02 SR1. It will be implemented with
N O T E
ChemStation B.04.01. The figures used in this chapater are from the G1315C DAD VL+. The
G1314E VWD figures will look the same.
Run Recovery Settings
For this recovery mode the CompactFlash Card must be in the detector.
C A U T I O N
If the LAN communiaction is interruppted, no data is stored.
The detector supports run buffering, which means that an amount of run
data (*.uv and *.ch files) is stored in a storage medium (CompactFlash
Card) in the detector until either it is overwritten or the detector
undergoes a power cycle.
If there is a temporary network failure or the PC is not able to constantly
take the data, the stored data is transferred to the ChemStation
automatically when the network connection is restored or the PC can take
the data, so that no loss of data occurs.
If there is a permanent network failure, the Run Recovery dialog box
allows you to restore the stored data to the data directory. From there
you can copy the files to the directory where the files are corrupted or
not complete.
On very large recovery files it may take a long time to restore it to the Agilent ChemStation.
N O T E
A sequence will be stopped in case of a network problem.
When during recovery an error "Method/Sequence stopped" appears, the instrument
N O T E
logbook shows an entry "No Run data available in device".
In this case refer to
80
Have the CompactFlash Card always inserted.
"No Run Data Available In Device"
on page 112.
Agilent 1200 Infinity Series VWD - User Manual