Troubleshooting and Diagnostics
Troubleshooting Based on Diagnostic Tests
Troubleshooting Based on Diagnostic Tests
If the actions described in the noise test procedures (see
Procedures"
actions should be performed by the service.
Electronic Noise Test
In case the electronic noise is higher than expected, see
Observed"
Stray Light Test
In case the intensity of the stray light/background noise is out of the expected
range, see
Failed Run of Standard Sample
In case the noise tests have passed, but the peak area for a standard is lower
than previous reference measurements, check the nebulizer (see
Nebulizer"
running the test using a backpressure loop instead of a column.
78
on page 71) do not solve the problems of failing tests, following
on page
80.
"No Signal is Observed"
on page 86), make sure you are using a fresh sample and consider
on page
79.