150
Signal including lambda information for the power meter (Auto).
Modulated signal to identify the fiber (at 270 Hz/330 Hz/1kH/2kHz).
Wavelengths activated one after the other (TwinTest mode).
Typical values, measured at 25°C unless specified.
Bi-directional testing
Wavelength
Insertion Loss
Dynamic range
IL uncertainty
IL repeatability
Display resolution
ORL
ORL range
ORL uncertainty
ORL repeatability
Display resolution
f
Length
Measurement range
Accuracy
- From 50 m to 20 km range
- Above 20 km
a. 1625nm not available with filtered version
b. side by side reference
c. without disconnection
d. from 10 to 45 dB range
e. from 20 to 40 dB
f. measurement @ 1550 nm with an index of refraction n = 1.468
For the OTDR specifications, please refer to

User Manual

770000102/08
1310 / 1490 / 1550 / 1625 nm
40 dB typ
+/- 0,25 dB
< 0,05 dB
0.01 dB
up to 55 dB
+/- 0,5 dB
< 0,1 dB
0.01 dB
150 km
+/- 30m
+/- 100m