Scatter coefficient
User
Auto
In Multimode, two predefined scatter coefficients are available:
The default values are given in the paragraph
Distance Unit
Define the unit of the distances displayed: km, kfeet, miles, meter, feet.
Results on trace
None
All
Graphics
If All or Graphics is selected, the reflectometry trace is displayed with a
dotted vertical line set on the end of launch cable
Cable is defined in the
of fiber
Otdr Connector Measurement
Detection
parameters
This parameter allows to choose if a measurement of the front connector
must be performed when acquisition is launched.
No
Yes
Splice
Select if a level of detection for splice must be defined.
User Manual
Selects for each wavelength, the backscatter coeffi-
cient of -99 dB to -50 dB by increments of 0.1dB.
Modification of the backscatter coefficient K changes
the measurements of reflectance and ORL.
Backscatter coefficients are selected automatically for
each wavelength.
Generic 50: 850 MM - > -66.3 dB
1300 MM -> -73.7 dB
Generic 62.5: 850 MM -> -66.1 dB
1300 MM -> -70.3 dB
the trace alone
the trace with results and markers.
the trace with markers only.
S
ETUP
.
In the results table, the first line corresponds to the
first event detected.
In the results, the first result corresponds to the front
connector measurement, at 0 meter (estimated
value).
770000102/08
Chapter 4 Configuring the reflectometry test
Configuring the test in OTDR Expert
"Reflectance" page
menu) and a dotted vertical line on the end
3.
(if the Launch
27