Enclosure (Touch) Leakage Current Test - GE B30 Technical Reference Manual

Patient monitor
Hide thumbs Also See for B30:
Table of Contents

Advertisement

B30 Patient Monitor

2.8 Enclosure (Touch) leakage current test

Perform this test to measure current leakage through exposed conductive surfaces on the
device under test during normal operation. Refer to the instructions contained with the safety
analyzer to perform enclosure leakage current test.
DMM
NOTE: The MD represents the network and voltage measuring instrument and its frequency
characteristics per IEC 60601-1.
(1)
Configure leakage tester as follows:
(2)
Power on device under test.
(3)
Read and record the current leakage indicated on tester.
NOTE: Center-tapped and non-center-tapped supply circuits produce different leakage
currents and the UL and IEC limits are different.
(4)
Change leakage tester switches to:
(5)
Read and record the current leakage indicated on the tester.
(6)
Change leakage tester switches to:
(7)
Read and record the current leakage indicated on the tester.
(8)
Change leakage tester switches to:
3-8
.
Document no
2044677-001
Leakage Tester
HIGH
LOW
GND
Open
Closed
0.15µF
1K
10
DMM set to measure AC voltage
Polarity - NORMAL
.
Neutral - OPEN
GND (Earth) - CLOSED
Polarity - NORMAL
Neutral - OPEN
GND (Earth) - CLOSED
Polarity - NORMAL
Neutral - OPEN
GND (Earth) - CLOSED
Polarity - NORMAL
Neutral - OPEN
NORM
Power Cord
RVS
GND
Probe to exposed conductive chassis
Device
Under
Test