13. Installation

Setting Air Drying Time

NOTE: This function is available on Rev. 7 controller boards with software revision SXXXR612 on
STATIM Classic, and software revision SXXXR700 on STATIM G4. However, this function is not
available on U.S. STATIM models.
For STATIM Classic, follow these steps:
1. Power up unit while pressing the STOP button to access the User Setup menu.
2. Using the UNWRAPPED and WRAPPED buttons, scroll to AIR DRYING and select it by
pressing the RUBBER AND PLASTIC button.
3. Use the UNWRAPPED and WRAPPED buttons to change the displayed values and the
RUBBER AND PLASTIC button to select and move to the next digit. Press the STOP button
when completed.
For STATIM G4, follow these steps:
1. Access the Service Menu as described in Using the LCD Service Menu on the STATIM G4 in
Chapter 3 and scroll to DRYING TIME.
2. Use the first pair of the arrow buttons to move from one digit to the next. Use the second pair
of arrow buttons to change digit values.
3. To save changes and return to the regular operating mode, press the STOP button.