2. Tools, Maintenance Schedules, Procedures and Testing
Recommended separate temperature and pressure meters for S unit calibration
temperature
Notes:
Pressure
Page 18
Fluke 51 Series II Digital Thermometer.
Fluke 80PK-26 SureGrip Tapered Temperature Probe.
The flexible thermocouple included with the Fluke digital thermometer will not be used for
calibrating SciCan sterilizers. If you wish to use the flexible thermocouple in the future,
SciCan advises you to consult the digital thermometer's manual for setting a temporary
offset, if required.
The above probe is 3.2mm (1/8") diameter.
Technicians already in possession of the Omega HH81A or equivalent, and the associated
probe can use this meter as a replacement for the Fluke 51 and 80PK-26 probe. Note that
the probes are of different diameter and a cassette compression fitting change may be
appropriate.
Druck DPI 705R Absolute Pressure Meter with external 0 - 7 bar absolute pressure
transducer with ¼" NPT female thread.