12. Printer and Data Logger
If MSD has: Corrupt or unreadable files and/or directories, refer to the STATIM screen to confirm
successful sterilization.
The MSD may have been unplugged while data was being written to it. The MSD should NOT be
unplugged until after "SAFELY REMOVE MSD <> MSD/FLASH DETECTED" is
displayed.
The corrupted files or directories may be lost. Reformat the MSD on your computer.
25 pin revision 2.x/5.x/6.x 2000 units only.
Unlike the revision 7.x unit, these units will not have a user or service menu and the LCD will NOT
display any messages that can be used for diagnostic purposes. The only way to tell that the
logger is not working correctly is either:
If either of these faults occurs, proceed as follows to ensure everything is setup correctly:
1. Check the cable connection, particularly the orientation.
2. Check the power connection.
3. Ensure the lower red LED is lit. It is located between the Data Logger's serial port and power
input.
4. Check that the Mass Storage Device (MSD) is properly inserted in the appropriate slot of the
Data Logger.
5. Repeat the instructions for Installing the Data Logger on your STATIM.
The Data Logger may be reset by unplugging its power supply, disconnecting the MSD and
waiting 10 seconds. Then re-connect the power adapter and insert the MSD into the Data Logger.
If the problem persists, contact the SciCan Service Center.
STAT IM 2000/2000S
STATIM 2000/2000S/2000 G4 Service Guide
Service Guide
96-108639 Rev 3.0
IM unit.
Page 255