5. STAT IM 2000S Calibration
10. During sterilization and after all necessary adjustments have been made, the measured
chamber pressure indicated by the reference meter and the STAT IM 2000S LCD should
not differ by more than 6 kPa.
11. A consecutive failure to achieve the above conditions after adjustment indicates that one
or more of the following components may be faulty:
Chamber thermocouple
Pressure sensor
Pressure interface board
Controller board (Thermocouple amplifier or A to D converter).
12. When the calibration is complete, press the STOP button on the control box and wait for
the beeps to signal that the cassette may be removed. Press the STOP button again to
reset the unit.
4. Steam generator thermocouple re-calibration
Repeat the steam generator thermocouple calibration procedure in section 2 described above.
5. Switch off the unit and remove all calibration equipment.
STAT IM 2000/2000S
STATIM 2000/2000S/2000 G4 Service Guide
Service Guide
96-108639 Rev 3.0
Page 107