SciCan STATIM 2000 Service Manual page 189

Hide thumbs Also See for STATIM 2000:
Table of Contents

Advertisement

8. Water Pumps, Reservoir, and Compressor
If the unit is being upgraded with the new style reservoir and the choice has been made to
upgrade the software to the latest version, the conductivity set-up in the service menu will need to
be modified.
Modifying the conductivity set up in the service menu
To modify the conductivity set up, proceed as follows:
1. Enter Service Menu.
2. Select "Conductivity Setup". (Pressing the Unwrapped button will allow you to scroll through
the menu).
3. When "Conductivity Setup" is on the top line of the LCD, press the Rubber or Plastic to enter
the sub menu. The display should be similar to the example below.
4. The L= value at this point SHOULD be a numerical value (e.g. 0.3 uS). Press the Wrapped
button to change the L= value to "FLOAT".
5. Press the Stop button. This will save and exit out of the Service Menu.
Removing and replacing the float sensor
To remove and replace the float sensor, proceed as follows (See Figure 22)
1. Cut the appropriate cable ties holding the float sensor wires in the wiring harness and
disconnect the connector from the controller board (see below). Release the wires from the
harness so that they are free to be removed from the unit with the sensor.
2. Remove nut from float sensor.
STAT IM 2000/2000S
STATIM 2000/2000S/2000 G4 Service Guide
CD=xx.xuS/NNN/y.y ppm
L=LL.L
H=HH.H
G=GG.G
Service Guide
96-108639 Rev 3.0
Float sensor
Figure 22
Page 187